COD Set-Up MD 200
The complete COD laboratory from a single source
- ST-COD sealed tubes ready for use
- Suppression of chloride interference up to 1000 mg/L (LR & MR) up to 10000 mg/L (HR)
- Mercury free tube tests,in absence of chloride interference
- 4 ranges:
- Low range: 3 - 150 mg/L, meets ISO 15705:2002
- Middle-low range: 15 - 300 mg/L, meets ISO 15705:2002
- Middle range: 20 - 1500 mg/L, meets ISO 15705:2002
- High range: 200 - 15000 mg/L
The Lovibond® COD Setups allow highly sensitive and precise water testing with minimum effort. They measure the ST-COD concentration by photometric detection employing a linear relationship between absorbance and concentration.
After adding the sample to a Lovibond® COD tube test (LR, MR according to ISO 15705:2002), it is heated in the reactor for two hours at 150 °C and then analysed in the photometer.
The COD Setups comprise the photometer, 25 tube tests for each of the two lower measuring ranges, a reactor for sample digestion and a vial stand.
After adding the sample to a Lovibond® COD tube test (LR, MR according to ISO 15705:2002), it is heated in the reactor for two hours at 150 °C and then analysed in the photometer.
The COD Setups comprise the photometer, 25 tube tests for each of the two lower measuring ranges, a reactor for sample digestion and a vial stand.
Part Number 2892602
Applications
- Raw Water Treatment
- Waste Water Treatment
Product information
Auto – OFF | Yes |
Portability | Benchtop |
Compliance | CE |
Testparameter | Measuring Range | Method No. | Chemical Method |
---|---|---|---|
COD HR TT | 200 - 15000 mg/L COD | M132 | Dichromate / H2SO4 |
COD LMR TT | 15 - 300 mg/L COD | M133 | Dichromate / H2SO4 |
COD LR TT | 3 - 150 mg/L COD | M130 | Dichromate / H2SO4 |
COD MR TT | 20 - 1500 mg/L COD | M131 | Dichromate / H2SO4 |
Title | Measuring Range | Chemical Method | Packaging Unit | Part Number | Add to quotation cart |
---|---|---|---|---|---|
COD LR/25 | 3 - 150 mg/L CODb) | Dichromate / H2SO4 | 25 pc. | 2420720 |
|
COD LR/25, mercury free | 3 - 150 mg/L CODb) | Dichromate / H2SO4 | 25 pc. | 2420710 |
|
COD LR/150 | 3 - 150 mg/L CODb) | Dichromate / H2SO4 | 150 pc. | 2420725 |
|
ValidCheck COD 40 mg/L + TOC 16 mg/L | 3 - 150 mg/L CODb) | Dichromate / H2SO4 | 1 pc. | 48371225 |
|
Validcheck COD 120 mg/L + TOC 48 mg/L | 3 - 150 mg/L CODb) | Dichromate / H2SO4 | 1 pc. | 48371425 |
|
ValidCheck WW Effluent Multistandard NH4-N/COD/TOC/NO3-N/PO4-P/TP | 3 - 150 mg/L CODb) | Dichromate / H2SO4 | 1 pc. | 48399612 |
|
COD MR/25 | 20 - 1500 mg/L CODb) | Dichromate / H2SO4 | 25 pc. | 2420721 |
|
COD MR/25, mercury free | 20 - 1500 mg/L CODb) | Dichromate / H2SO4 | 25 pc. | 2420711 |
|
COD MR/150 | 20 - 1500 mg/L CODb) | Dichromate / H2SO4 | 150 pc. | 2420726 |
|
COD MR/150, mercury free | 20 - 1500 mg/L CODb) | Dichromate / H2SO4 | 150 pc. | 2420716 |
|
ValidCheck COD 500 mg/L + TOC 200 mg/L | 20 - 1500 mg/L CODb) | Dichromate / H2SO4 | 1 pc. | 48371625 |
|
ValidCheck WW Influent Multistandard NH4-N/COD/TOC/NO3-N/PO4-P/TP | 20 - 1500 mg/L CODb) | Dichromate / H2SO4 | 1 pc. | 48399712 |
|
COD HR/25 | 200 - 15000 mg/L CODb) | Dichromate / H2SO4 | 25 pc. | 2420722 |
|
COD HR/25, mercury free | 200 - 15000 mg/L CODb) | Dichromate / H2SO4 | 25 pc. | 2420712 |
|
COD HR/150 | 200 - 15000 mg/L CODb) | Dichromate / H2SO4 | 150 pc. | 2420727 |
|
ValidcCheck COD 5000 mg/L + TOC 2002 mg/L | 200 - 15000 mg/L CODb) | Dichromate / H2SO4 | 1 pc. | 48371825 |
|
COD LMR/25 | 15 - 300 mg/L CODb) | Dichromate / H2SO4 | 25 pc. | 2423120 |
|
Title | Part Number | Add to quotation cart |
---|---|---|
Thermoreactor RD 125 | 2418940 |
|
Pipette, 200 µl | 365042 |
|
Pipette Tips | 365032 |
|
Service plan - 3 years for MD100/MD110/MD200 | 19802801 |
|
Fixed price service package for MD100/MD110/MD200 | 19802701 |
|
Cuvette stand for 10 round cuvettes Ø 16 mm | 418957 |
|
Automatic pipette, 1-5 ml | 419076 |
|
Pipette tips, 1-5 ml (white) 100 pc. | 419066 |
|
Automatic pipette, 0,1-1 ml | 419077 |
|
Pipette tips, 0,1-1 ml (white), 1000 pc. | 419073 |
|
Adapter for round cuvettes 13 mm Ø | 19802192 |
|
Adapter for round cuvettes 16 mm ø | 19802190 |
|
Standard Solution COD 100 mg/l | 2420803 |
|
Standard Solution COD 500 mg/l | 2420804 |
|
Standard Solution COD 5000 mg/l | 2420805 |
|
Factory calibration certificate ISO 9001 for MD100/MD110/MD200 | 999750 |
|
- Photometer
- COD Reactor RD 125
- COD tube tests (0 - 150 mg/l and 0 - 1500 mg/l)
- Tube stand
- Instruction manual